DUI Videos Chemical Interferent

Video 08

Limonene24 - Low and High Cal. Checks - Lowering 0 - Contamination of Cal Checks

Intoxilyzer 5000 in need of maintenance. 100 alcohol standard at 34 C. Exposure to mixture of room temperature 50 alcohol standard and GOJO containing limonene bubbled into breath tube during air blanks plus GOJO on hands near flask air entry. After initi

Video 09

Limonene30 - Low & High Cal. Checks Caused by Limonene

Intoxilyzer 5000 in need of maintenance. Experiments with GOJO hand cleaner containing limonene partially dissolved in cold 50 mg/100mls alcohol standard bubbled into breath tube during some air blanks. Effects on cal. checks with 200 alcohol standard. So

Video 10

Limonene31 - High Cal. Checks Caused by Limonene

Intoxilyzer 5000 in need of maintenance. 100 alcohol standard at 34C. Note high cal. checks on ACACA and ACABA sequences. Calibration check system (sample chamber, interferent detect/subtract/add circuitry, tubing, simulator, or standard) has been comprom

Video 11

Limonene25 - High Cal. Checks - Limonene

Intoxilyzer 5000 in need of maintenance. 100 alcohol standard at 34 C. Prior exposure to mixture of room temperature 50 alcohol standard and GOJO containing limonene bubbled into breath tube during air blanks plus GOJO on hands near flask air entry. ACACA

Video 15

Limonene1 Cal. Checks 103, 81, 96, 85, 111, 105

Experiment to explore the effects of a hand cleaner such as GOJO containing limonene in the ambient air respecting reliability of the alcohol standard. 100 SAS in simulator at 34.0C. Operator runs ACACA mode. Cal. check drops from 103 to 81, 96, 85 when h

Video 16

ProcessorErrors 0030d Ambient Threshold Varies with Ambient Conditions

Main simulator has 100 SAS at 34.0C. This video is a sequel to two longer videos. Experiment using cold extra simulators at 40 mgs/100mls and 100 mgs/100mls gradually warmed to simulate slow changes in ambient ethanol in room air. Instrument does not trig

Video 49

Intoxilyzer 5000 64 series interferents

Intoxilyzer 5000 64 series interferents

Video 52

Cell Phone RFI Error Interferent 64 Series

Cell Phone RFI Error Interferent 64 Series

Video 53

Cell Phone RFI Error Interferent 64 Series 0001

Cell Phone RFI Error Interferent 64 Series 0001